handheld-oxygen-analyzers-for-medical-gases-aii-2000-and-aii-2000-palm-o2.png
Analytical Industries’ AII-2000 series or Palm O2 handheld oxygen analyzers provide a very easy way to verify oxygen and other medical gases used in medical or veterinary clinics. These FDA approved inexpensive medical gas analyzers use maintenance-free advanced galvanic oxygen sensors and will operate for up to 32 months. Based on decades of experience in the design and manufacture of oxygen instrumentation for respiratory medical devices, both models have been specifically designed for spot checking oxygen levels in anaesthesia, intensive care, incubators or other variants of respiratory or oxygen therapy.

 

Features of Handheld Oxygen Analyzers for Medical Gases

  • Very easy to operate
  • Maintenance free oxygen Sensors
  • 0-100% O2 range
  • Up to 32 months sensor lifetime
  • Integral or remote sensor options
  • Long battery life: up to 18 months of continuous operation for AII-2000 Palm O2

 

 

Technical Specifications

  AII-2000 A (and AHC) AII-2000 M AII-2000 Palm R AII-2000 Palm O2
Measurement range 0-100% oxygen
Accuracy < ±2% of range at constant conditions
Response time T90 < 10 seconds
Sensitivity (LDL) 0.1%
Linearity < 2% of scale
Sensor AII-11-60 (A model) or AII-11-60-HC (HC model) AII-11-60 AII-11-75-PO2R
Sensor life at 25ºC (77°F) and 1 atm 60 months in air 32 months in air
Calibration interval After 8 hours of use
Inlet pressure Ambient 
0.34-2 barg (5-30 psig) with atmospheric vent
Flow rate 0.2-10 l/m (0.4-20 SCFH)
Gas connections 1x16 mm thread or push-in flow diverter with o-ring seal
Display 3½ digit backlit LCD 7 x 3.5cm (2.75 x 1.375"); resolution 0.1 3 digit LCD 2.8 x 1.6cm (1.1 x 0.625"); resolution 0.1
Dimensions 9.1 x 14.9 x4cm (3.6 x 5.9 x 1.6") 6.9 x 10.4 x 3.4cm (2.72" x 4.1" x 1.35")
Weight 280g (10oz) 196g (7oz)
Compensation Temperature
Alarms None User configurable 
HI & LOW alarms
None
LED indicators None Alarm status None
Storage temperature 0° to 45°C (32°F to 113°F) on intermittent basis -20° to 60°C (-4°F to 140°F) on intermittent basis
Operating temperature 5° to 45°C (41°F to 113°F)
Humidity Non-condensing 0-95% RH
Power 2 AA Alkaline batteries; 
1,200 hours continuous use
2 AA alkaline batteries; 13,000 hours continuous use